GE water pump 800 Manual

PDF Manuals
16 Pages English

GE water pump 800
GE Analytical Instruments, Inc. ©2005H Page 8 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
CAUTION:
IF THERE ARE ANY VARIATIONS, CONSTANTS MUST BE ENTERED THROUGH
THE FACTORY ACCESS MENU. ACCESS TO THIS MENU IS LIMITED AND WILL
REQUIRE A PASSWORD. ENSURE THAT ACCESS TO THIS MENU IS
RESTRICTED. ANY WRONGFUL DATA ENTRY WILL RESULT IN THE ANALYZER
PERFORMING OUT OF SPECIFICATION!!!
5.6.1.1. From the MAIN menu select SETUP, press Enter.
5.6.1.2. From the SETUP menu, select CALIBRATE, press Enter.
5.6.1.3. Select TC ZERO OFFSET, press Enter.
TC ZERO CORRECTION
-1.3250000
ENTE
R
CLEA
R
ENT = SAVE CHAR/NUM
CLR = RESTART/EXIT
A
RROWS = CHANGE CHA
R
FIGURE SEVEN: TOC ZERO OFFSET Screen
5.6.1.4. Enter 911 by using the arrows, and then press Enter.
NOTE:
ENTERING 911 WILL IN NO WAY AFFECT THE VALUE OF THE TC ZERO
CORRECTION.
FACTORY ACCESS V. 0.00
ERASEERRS
ERASE CARD
ENTER
CLEA
R
PRNT SETUP
SETEEPROM
COND ZERO
A
NALOG #
FACTMODE
S
RAW DAT
A
SET 4-20M
A
FIGURE EIGHT: FACTORY ACCESS Menu
5.6.1.5. Using the arrows, select SETEEPROM, and press Enter.
5.6.1.6. Constants will be displayed in the same order as the printed or
recorded constants list.
5.6.1.7. Once scrolling through the list has begun, the entire list must be
seen. DO NOT PRESS THE CLEAR KEY. Pressing the Clear key will
result in the erasing of the Constant Value entered. To move up
or down the list use the arrow keys.
0.800600
C1 CONST:
ENT = CHANGE VALUE
ENTE
R
CLEA
R
CLR = SET TO DEFAULT
A
RROWS = NEXT VALUE
Related Manuals for GE water pump 800